Introduction -- The instrument in
room 109 is a Spectrametrics DC-Plasma Emission Spectrometer (DCP)
This unit uses a DC-plasma as its heat source and an echelle
grating for high resolution. The spectometer is capable of determining up
to 20 elements in solution simultaneously. Elements with a significant
emission line between 200 nm and 700 nm can be analyzed. The working
detection limits for most elements is 0.02 mg/L.
History -- In 1969, Spectrametrics
introduced the first commercially available direct current plasma system
which used a right angle plasma as an excitation source and a crossed
prism echelle spectrometer as the analytical detector. This system, known
as the SpectraJet I, suffered from poor light though put as well as a
substantial interference of the analyte signal by the plasma background
column due to the right angle nature of the arc geometry.
In 1974, Spectrametrics replaced the
right-angle plasma with an inverted "V" configuration plasma
known as the SpectraJet II. The DC argon plasma is initiated by arcing a
high-voltage spark from cathode to anode. Once ignited, the plasma is
sustained by a relatively low voltage, about 50 volts with a current of
about seven ampers. Samples are nebulized and introduced into the
excitation area in aerosol form. Supplementary argon flows around the
cathode and anode to maintain a stable discharge. It is important to note
that when this source is used with the echelle system, only light from the
excitation region is focused into the spectrometer. The intense continuum
from the plasma column and the plume is not observed.
In 1977, Spectrametrics introduced the
SpectraJet III. In this three-electrode direct current argon plasma the
plasma jet is formed between two spectrographic carbon anodes and a
tungsten cathode in an inverted
"Y" configuration. This design resulted in improved stability
and better detection limits than previously reported for dc plasmas. Of
particular importance is its stability in the presence of varying solvent
types such as those containing large amounts of dissolved solids,
organics, or high acid/alkaline concentrations. The sample excitation
region and photometric observation area of the SpectraJet III is centered
uniquely in the crook of the "Y "where spectral contribution
from the plasma continuum is minimal. This zone reaches an excitation
temperature in excess of 8000 degrees centigrade and is evident in the
above photograph where strontium (reddish) is
being excited in an area distinct from the plasma continuum. The plasma
requires less than 1000 watts of power and consumes approximately 8 L/min
of welder's grade argon.
Detectors -- The Spectraspan III, V, and
VII echelle spectrometers are able to operate in either the sequential or
simultaneous mode. The spectral output of these spectrometers consists of
approximately 100 spectral orders compressed into a 4 inch by 5 inch (10
cm by 12 cm) area. A kinematically mounted cassette is located at the
focal plane of these spectrometers. The cassette is a 6 inch square by 1
inch thick aluminum assembly which has a locating pin on one side. The
exit slits are located under a protective cover. In the sequential mode,
the cassett e
contains one exit slit and the prism/grating assembly is scanned to allow
the appropriate wavelength to hit the exit the slit. In the simultaneous
mode, the prism/grating assembly is fixed and the cassette contains up to
24 exit slits for the selected emission lines. This ability to be replaced
allows the laboratory to have a cassette for each application so that the
best emission line for each element can be used. Behind each exit slit, a
periscope is used to redirect the emission light to a photomultiplier tube
where the quantity of light is detected and the concentration of the
associated element is quantitated. The mirrors used in these periscopes
have been optimized for 45 degree incidence angle and have better than 91
percent reflectivity at 190 nm and nearly 95 percent in the visible region
of the spectrum.
Grating -- The basis of the method is the
measurement of atomic emission by an optical spectroscopic technique.
Samples are nebulized and the aerosol that is produced is transported to
the plasma where excitation occurs.

Characteristic atomic-line emission spectra are produced by a direct
current plasma (DCP). Spectra are produced by an echelle grating
spectrometer and the intensities of the lines are moitored by film or
photomultiplier tubes. The photocurrents from the photomultiplier tubes
are processed and controlled by a computer system. A background correction
technique may be required to compensate for variable background
contribution to the determination of trace elements. Background must be
measured adjacent to analyte lines on samples during analysis. The
position selected for the background intensity measurement, on either or
both s ides
of the analytical line, will be determined by the spectrum adjacent to the
analyte line. The position used must be free of spectral interference and
reflect the same change in background intensity as occurs at the analyte
wavelength measured. Background correction is not required in cases of
line broadening where a background correction measurement would actually
degrade the analytical result.
Physical interferences are generally
considered to be effects associated with the sample nebulization and
transport processes. Such properties as change in viscosity and surface
tension can cause significant inaccuracies especially in samples which may
contain high dissolved solids and/or acid concentrations. The use of a
peristaltic pump may lessen these interferences. If these types of
interferences are operative, they must be reduced by dilution of the
sample and/or utilization of standard addition techniques. Also, it has
been reported that better control of the argon flow rate improves
instrument performance. This is accomplished with the use of mass flow
controllers.
Chemical interferences are
characterized by molecular compound formation, ionization effects and
solute vaporization effects. Normally these effects are not pronounced
with the DCP technique, however, if observed they can be minimized by
careful selection of operating conditions (that is, observation position,
and so forth), by buffering of the sample, by matrix matching, and by
standard addition procedures. These types of interferences can be highly
dependent on matrix type and the specific analyte element.

Elements such as lithium, sodium, potassium, rubidium, cesium, calcium,
magnesium, etc. can cause enhancement of the net signl-to-noise ratio for
many of the elements in this method. This effect can be both controlled
and utilized advantageously by the addition of lithium, or cesium at a
final concentration of 1,000 mg/L, to the blank, the standards, and the
samples.
Echellogram -- The spectral output of a crossed dispersed echelle
spectrometers consists of about 100 spectral orders compressed into a 4
inch by 5 inch (10 cm by 12 cm) area. The vacuum ultraviolet light (below
180 nanometers) occurs in orders 124 and above and is seen at the top of
the spectral window while the infrared light (above 850 nm) begins in
order 23 at the lower left. Visible and ultraviolet light which is used
for most atomic elemental analysis, occurs between these two extremes. If
laid end to end, the useful segments of the echellogram would be
equivalent to a spectrum 19 feet long (6 meters). The resolution of an
echelle spectrometer is typically 10 to 16 times that of a conventional
spectrometer of equivalent focal length. In this particular case, the
spectral bandwidth is 0.0038 mn. and reciprocal linear dispersion is 0.06
nm/mm in the 112th order 
The two bright spots in the top center are the dual emission images of the
253.6519 nm
mercury registration lines while the bright blue-white streaks nearer the
bottom are caused by the black body emission of the source and occur at
approximately 400 nm. Each elemental
emission image is approximately 50 microns wide and 300 to 500 microns
high, however they
can be broadened by spectral interference or temperatur
e and doppler effects.
|