Themes > Science > Chemistry > Analytical Chemistry > Methods and Instrumentation > Direct-Current Plasma Excitation Source

Introduction -- The instrument in room 109 is a Spectrametrics DC-Plasma Emission Spectrometer (DCP)   This unit uses a DC-plasma as its heat source and an echelle grating for high resolution. The spectometer is capable of determining up to 20 elements in solution simultaneously. Elements with a significant emission line between 200 nm and 700 nm can be analyzed. The working detection limits for most elements is 0.02 mg/L.

History -- In 1969, Spectrametrics introduced the first commercially available direct current plasma system which used a right angle plasma as an excitation source and a crossed prism echelle spectrometer as the analytical detector. This system, known as the SpectraJet I, suffered from poor light though put as well as a substantial interference of the analyte signal by the plasma background column due to the right angle nature of the arc geometry.

In 1974, Spectrametrics replaced the right-angle plasma with an inverted "V" configuration plasma known as the SpectraJet II. The DC argon plasma is initiated by arcing a high-voltage spark from cathode to anode. Once ignited, the plasma is sustained by a relatively low voltage, about 50 volts with a current of about seven ampers. Samples are nebulized and introduced into the excitation area in aerosol form. Supplementary argon flows around the cathode and anode to maintain a stable discharge. It is important to note that when this source is used with the echelle system, only light from the excitation region is focused into the spectrometer. The intense continuum from the plasma column and the plume is not observed.

In 1977, Spectrametrics introduced the SpectraJet III. In this three-electrode direct current argon plasma the plasma jet is formed between two spectrographic carbon anodes and a tungsten cathode in an inverted "Y" configuration. This design resulted in improved stability and better detection limits than previously reported for dc plasmas. Of particular importance is its stability in the presence of varying solvent types such as those containing large amounts of dissolved solids, organics, or high acid/alkaline concentrations. The sample excitation region and photometric observation area of the SpectraJet III is centered uniquely in the crook of the "Y "where spectral contribution from the plasma continuum is minimal. This zone reaches an excitation temperature in excess of 8000 degrees centigrade and is evident in the above photograph where strontium (reddish) is being excited in an area distinct from the plasma continuum. The plasma requires less than 1000 watts of power and consumes approximately 8 L/min of welder's grade argon.

Detectors -- The Spectraspan III, V, and VII echelle spectrometers are able to operate in either the sequential or simultaneous mode. The spectral output of these spectrometers consists of approximately 100 spectral orders compressed into a 4 inch by 5 inch (10 cm by 12 cm) area. A kinematically mounted cassette is located at the focal plane of these spectrometers. The cassette is a 6 inch square by 1 inch thick aluminum assembly which has a locating pin on one side. The exit slits are located under a protective cover. In the sequential mode, the cassette contains one exit slit and the prism/grating assembly is scanned to allow the appropriate wavelength to hit the exit the slit. In the simultaneous mode, the prism/grating assembly is fixed and the cassette contains up to 24 exit slits for the selected emission lines. This ability to be replaced allows the laboratory to have a cassette for each application so that the best emission line for each element can be used. Behind each exit slit, a periscope is used to redirect the emission light to a photomultiplier tube where the quantity of light is detected and the concentration of the associated element is quantitated. The mirrors used in these periscopes have been optimized for 45 degree incidence angle and have better than 91 percent reflectivity at 190 nm and nearly 95 percent in the visible region of the spectrum.

Grating -- The basis of the method is the measurement of atomic emission by an optical spectroscopic technique. Samples are nebulized and the aerosol that is produced is transported to the plasma where excitation occurs. 

Echelle Grating

Characteristic atomic-line emission spectra are produced by a direct current plasma (DCP). Spectra are produced by an echelle grating spectrometer and the intensities of the lines are moitored by film or photomultiplier tubes. The photocurrents from the photomultiplier tubes are processed and controlled by a computer system. A background correction technique may be required to compensate for variable background contribution to the determination of trace elements. Background must be measured adjacent to analyte lines on samples during analysis. The position selected for the background intensity measurement, on either or both sides of the analytical line, will be determined by the spectrum adjacent to the analyte line. The position used must be free of spectral interference and reflect the same change in background intensity as occurs at the analyte wavelength measured. Background correction is not required in cases of line broadening where a background correction measurement would actually degrade the analytical result.

Physical interferences are generally considered to be effects associated with the sample nebulization and transport processes. Such properties as change in viscosity and surface tension can cause significant inaccuracies especially in samples which may contain high dissolved solids and/or acid concentrations. The use of a peristaltic pump may lessen these interferences. If these types of interferences are operative, they must be reduced by dilution of the sample and/or utilization of standard addition techniques. Also, it has been reported that better control of the argon flow rate improves instrument performance. This is accomplished with the use of mass flow controllers.

Chemical interferences are characterized by molecular compound formation, ionization effects and solute vaporization effects. Normally these effects are not pronounced with the DCP technique, however, if observed they can be minimized by careful selection of operating conditions (that is, observation position, and so forth), by buffering of the sample, by matrix matching, and by standard addition procedures. These types of interferences can be highly dependent on matrix type and the specific analyte element.


Elements such as lithium, sodium, potassium, rubidium, cesium, calcium, magnesium, etc. can cause enhancement of the net signl-to-noise ratio for many of the elements in this method. This effect can be both controlled and utilized advantageously by the addition of lithium, or cesium at a final concentration of 1,000 mg/L, to the blank, the standards, and the samples.


Echellogram -- The spectral output of a crossed dispersed echelle spectrometers consists of about 100 spectral orders compressed into a 4 inch by 5 inch (10 cm by 12 cm) area. The vacuum ultraviolet light (below 180 nanometers) occurs in orders 124 and above and is seen at the top of the spectral window while the infrared light (above 850 nm) begins in order 23 at the lower left. Visible and ultraviolet light which is used for most atomic elemental analysis, occurs between these two extremes. If laid end to end, the useful segments of the echellogram would be equivalent to a spectrum 19 feet long (6 meters). The resolution of an echelle spectrometer is typically 10 to 16 times that of a conventional spectrometer of equivalent focal length. In this particular case, the spectral bandwidth is 0.0038 mn. and reciprocal linear dispersion is 0.06 nm/mm in the 112th order


























The two bright spots in the top center are the dual emission images of the 253.6519 nm
mercury registration lines while the bright blue-white streaks nearer the bottom are caused by the black body emission of the source and occur at approximately 400 nm. Each elemental
emission image is approximately 50 microns wide and 300 to 500 microns high, however they
can be broadened by spectral interference or temperatur e and doppler effects.


Information provided by: http://www.chemistry.nmsu.edu